--- In AVR-Chat@yahoogroups.com, "kernels_nz" <kernels@...> wrote: > Wow, I see to be generating a bit of interest ... Although I share Roy's distaste for allowing the wrong skill set to control a design approach, I can see reasons why what Hein describes could be optimum. I think we're looking not only to resolve 10,000 steps in 15 mV, but to have comparable total accuracy also. Or, if we're going to calibrate, at least have sources of inaccuracy that are constant and stable in between calibrations. So, to amplify the signal, and thus allow a less accurate A-to-D to be used, we'd need around 0.01% accuracy in the amplifier. I'm not saying that can't be done, but that's not the question. Can it be done more cheaply that using a mass-market delta-sigma A-to-D converter? The accuracy of these converters is pretty much guaranteed by the conversion technique. Using the load cell excitation voltage as the reference and having no analog gain eliminates a whole raft of error sources. I think this is the right approach. Also, I think the thermocouple effects are a red herring. Indeed, you can't mitigate with low impedance because the source is low impedance. But it is normal to use wiring with the same material on the signal and ground side (or the two sides of a differential connection), not the differing materials of a thermocouple junction. You're also not going to use materials chosen to produce the maximum thermoelectric effect. Probably the only part of the current path that isn't copper will be the load cell itself. As long as both connections to that are at the same temperature, there should be no problem. Graham.
Message
Re: Weigh - Scale Firmware
2007-10-11 by Graham Davies
Attachments
- No local attachments were found for this message.