Re: [Doepfer_a100] Re: A-110 VCO calibration

2005-08-10 by Florian Anwander

Hi


>>The full set-up procedure is in the A-110 Service Manual, which is 
>>available at the Doepfer website (there's a link on the A-110 page).
> Hmm... thanks but not very helpful, I do not wish to carry 
> out such a procedure, as I may make the matter worse with
> the VCO.
Then you heave to bring it to a shop, who gives service for Doepfer.

Nevertheless it sould be quite easy:
- switch off the A100 system
- remove the screws from the A110
- do NOT disconnect the power of the A110,
- cover the soldering side of the A110 with a piece of paper; you
   may fix it simply with an adhesive tape (This prevents unwanted
   shortages, though it is quite diffcult to destroy really
   something)
- switch on the A100 and let it warm up a quarter of an hour.
- set the "Tune" poti to center position.
- take a small screw driver and tune the VCO with the trimmer "P6"
   (it is placed direct aside the writing "DOEPFER A110" on the pcb).
- DO NOT TOUCH ANY OTHER TRIMMER!
- switch off the A100
- remove the shielding paper on the A110's back
- fix the A110 with the screws in the frame

Best regards, Florian