Re: [emax] Emu emax1 with problem in the input sampler

2016-01-25 by Matt Ferrari

You might not have the threshold level set low enough to detect a signal
on the emax II its choice number 5 on the sample menu
also there is a gain choice #1 and an arm sample, force sample choice

these should be all the choices involved in recording a sample


----- Original Message ----- 
From: Ted Summers djtbs1@... [emax]
To: emax@yahoogroups.com
Sent: Sunday, January 24, 2016 4:08 PM
Subject: Re: [emax] Emu emax1 with problem in the input sampler



Did you go through the manuals / guides and do:
calibration
Emax Clinic
Diagnostics

Is that the only feature not working, or perhaps more than that?

-T

On Jan 24, 2016, at 12:59 AM, yusteromerooscar@... [emax] 
<emax@yahoogroups.com> wrote:

Recently I bought a Emu emax1 and I have a problem with the input sampler, 
it is not detecting input signal and (of course) no record... any clue of 
what is happening??

Thank you very much.