> Hi list, > > I have been taking some detailed measurements from my A-127 (for > comparison against a heavily-modified Plan B M12) and have > noticed that one of the three sections is adjusted quite > differently from the other two, which agree very closely. Thus > I'm not sure which sections are adjusted as they would have been > when I originally bought it (it's likely I have 'fiddled with it' > over the years!), so if anyone has the service manual and can > detail or send me the calibration procedure I would very much > appreciate it! > > Many thanks, > Tim There are four trimming potentiometers on each A-127 board: P6/100k: frequency offset (defines the absolute filter frequency when the VCF Frq. control is fully CCW) P7/2k5 : frequency scale (defines the sensitivity of the CV input, can be adjusted e.g. to approximately 1V/oct) P8/100k: OTA offset 1 (adjusted for minimal CV feedtrough) P9/100k: OTA offset 2 (adjusted for minimal CV feedtrough) This is the order of the trimming potentiometers from left to right on the pcb (front view): P7, P6, P8, P9 Hope this helps Best wishes Dieter Doepfer