So measure tool reports the readings in the "random" order of the print out and you need to sort the results before you give them to QTR-Create-ICC. I do this in Excel. Don't sort by the actual Lab reading, sort by the expected value. That way if you really do have some out of order readings, QTR-Create-ICC will warn you. Mike 2008/10/10 slcphoto73 <slchapin@...>: > I have just printed Step-51-random.tif, run MeasureTool and measured > the three strips as strip-without-gaps. When I put the saved .txt file > into QTR-Create-ICC it tells me the Lab values are not in order and it > can't make the profile. And yes, I observe some Lab values out of > order in the "out" text file. > > I have repeated the measurement multiple times. I tried measuring as > patches but I couldn't properly center the spectrophotometer on the > patches, so it had the same problem only worse. > > What might be wrong and how do I correct it? > > I am running QTR v. 2.6.1.0, with CreateICC dated 9/18/2007. > > Thank you, > > - susan > >