Re: [QuadtoneRIP] Help using QTR-Create-ICC.exe

2008-10-10 by Michael King

So measure tool reports the readings in the "random" order of the
print out and you need to sort the results before you give them to
QTR-Create-ICC. I do this in Excel. Don't sort by the actual Lab
reading, sort by the expected value. That way if you really do have
some out of order readings, QTR-Create-ICC  will warn you.

Mike

2008/10/10 slcphoto73 <slchapin@...>:
> I have just printed Step-51-random.tif, run MeasureTool and measured
> the three strips as strip-without-gaps. When I put the saved .txt file
> into QTR-Create-ICC it tells me the Lab values are not in order and it
> can't make the profile. And yes, I observe some Lab values out of
> order in the "out" text file.
>
> I have repeated the measurement multiple times. I tried measuring as
> patches but I couldn't properly center the spectrophotometer on the
> patches, so it had the same problem only worse.
>
> What might be wrong and how do I correct it?
>
> I am running QTR v. 2.6.1.0, with CreateICC dated 9/18/2007.
>
> Thank you,
>
> - susan
>
>